Analytical results of nutrient ions by CFA
For nitrate, nitrite and phosphate ions, five recovery tests were carried out as follows: the seawater samples were diluted to the respective content specified in “Preparation of measuring solutions for CFA” and known amount of the nutrient ion standard solutions were added. The measuring solutions before and after addition were measured by the CFA bracketing method. The differences, the experimental addition mass fractions, were compared to the addition mass fractions in the gravimetric preparation. The results are shown in Table 2. For nitrate ions, except for the analysis of the lowest addition mass fraction, it was confirmed that for each artificial sample nearly 100% nitrate ions were recovered almost within the expanded uncertainty with a coverage factor of 2, U(k = 2), derived from the uncertainty of the bracketing method including the repeatability. Although the tests for nitrate ion of the lowest mass fraction, nitrite and phosphate ions showed deviations beyond the repeatability due to the low absolute mass fractions to be measured, it was determined that they were acceptable considering the uncertainty required for oceanographic observations. They were included in the uncertainty of the analytical results as matrix difference between sample and standard solutions, as shown in the previous report [12].
Table 2
Results of recovery tests for nutrient ions by CFA method.
| Nutrient | Base Seawater | Seawater containing rate | Added wpre / mg kg− 1 | Found *4 wdet / mg kg− 1 | Recovery*4 wdet / wpre |
| NO3− | HSW | 0.04*1 | 0.1046 | 0.1046 ± 0.0004 | 100.0% ± 0.4% |
| | MSW | 0.50*2 | 0.0459 | 0.0458 ± 0.0002 | 99.9% ± 0.5% |
| | ELSW | 0.75*3 | 0.00010 | 0.00008 ± 0.00001 | 75.2% ± 16.8% |
| NO2− | MSW | 0.75 | 0.00159 | 0.00164 ± 0.00002 | 102.7% ± 1.3% |
| PO43− | HSW | 0.75 | 0.0208 | 0.0202 ± 0.0002 | 97.3% ± 0.8% |
| *1 Corresponding to HSW analysis. *2 Corresponding to MSW analysis. *3 Corresponding to ELSW analysis. *4 Each figure following ± indicates the expanded uncertainty due to bracketing method including the repeatability (k = 2). |
Based on the discussion, three kinds of seawater samples (candidate NMIJ CRMs) were analyzed by CFA, whose results were shown in Table 3.
Table 3
Measured values of each candidate NMIJ CRM by CFA method *1
| | Mass fraction for HSW / mg kg− 1 | Mass fraction for MSW / mg kg− 1 | Mass fraction for ELSW / mg kg− 1 |
| NO2- | 0.00123 ± 0.00013 (n = 2) | 0.0183 ± 0.0007 (n = 2) | 0.00165 ± 0.00012 (n = 2) |
| NO3- | 2.7528 ± 0.0071 (n = 4) | 0.9565 ± 0.0020 (n = 4) | 0.0013 ± 0.0003 (n = 2) |
| PO43- | 0.2876 ± 0.0010 (n = 2) | 0.1013 ± 0.0010 (n = 2) | 0.0015 ± 0.0002 (n = 2) |
| *1 Each figure following ± indicates the combined standard uncertainty. The combined standard uncertainty uc was calculated from the following equation \({u}_{c}=\sqrt{{\left({s}_{\text{r}}/\sqrt{n}\right)}^{2}+{u}_{\text{mtrx}}^{2}+{u}_{\text{std}}^{2}}\), where n is a number of replication of measurements using ten bottles, sr is repeatability SD obtained by analysis of variance from measured values for ten bottles × n, umtrx is the uncertainty due to matrix difference between the sample and standard solution sets (refer to the previous report [12]) and ustd is the standard uncertainty due to the standard solution of the analyte. |
Analytical Results Of Nitrate Ions By Ic
The linearity of each calibration curve for nitrate ions in both UV absorbance detection and electric conductivity detection was verified since the coefficient of correlation of the calibration curve for each detection was more than 0.9999 in the mass fraction range between 0.8 mg kg− 1 and 8.3 mg kg− 1 with dilution to 0.95 g g− 1 seawater. Therefore, the mass fractions of nitrate ions in both HSW and MSW were measured by the standard addition method.
In the case of the AS10 column with an HCl eluent, the mass fractions of nitrate ions in both HSW and MSW were precisely measured with UV absorbance detection (Fig. 2), although they could not be measured with electric conductivity detection because of a huge vicinity peak of chloride ions and an unstable baseline. Although nitrate ions could be detected more sensitively at the wavelength of 210 nm, the wavelength of 225 nm was chosen from the viewpoint of avoiding interference of bromide ions because the molar absorption coefficient of bromide ions in longer wavelength became rapidly lower than that of nitrate ions [21]. On the other hand, when bromide ions were efficiently removed by the desalination procedure for chloride, bromide and iodide ions, the more effective wavelength of 210 nm could be used with no interference of bromide ions. Although the detection at 210 nm with the desalination was advantageous in the sensitivity of nitrate ions compared to the detection at 225 nm with no desalination, the uncertainty of the quantitative result was not improved because of the addition of tedious procedure. It should be noted that the quantitative value of the detection at 225 nm with no desalination agreed with that of the detection at 210 nm with desalination (Table 4). This fact partly supported no interference from any coexisting ions.
Table 4
Measured mass fractions of nitrate ions in candidate NMIJ CRMs by IC method
| | Pretreatment *1 | Column / Eluent | Detection | HSW *2 | MSW *2 | ELSW *2 |
| A | only dilution | AS10 / HCl | UV absorbance (225 nm) | 2.748 ± 0.010 (n = 5) | 0.934 ± 0.005 (n = 4) | DL = 0.005 *3 |
| B | only dilution | AS23 / carbonate | UV absorbance (210 nm) | 2.748 ± 0.021 (n = 1) *5 | 0.937 ± 0.009 (n = 1) *6 | - |
| C | only dilution | AS23 / carbonate | electric conductivity | 2.725 ± 0.031 (n = 1) *5, *7 | 0.936 ± 0.015 (n = 1) *6, *7 | - |
| D | desalination-1 | AS10 / HCl | UV absorbance (210 nm) | 2.751 ± 0.026 (n = 2) | 0.936 ± 0.028 (n = 1) | - |
| E | desalination-1 | AS12A / carbonate | UV absorbance (210 nm) | 2.724 ± 0.032 (n = 2) | 0.932 ± 0.022 (n = 1) | - |
| F | desalination-1 | AS12A / carbonate | electric conductivity | 2.759 ± 0.024 (n = 1) | 0.931 ± 0.012 (n = 1) | - |
| G | desalination-2 | AS12A / carbonate | UV absorbance (210 nm) | 2.738 ± 0.026 (n = 1) *4, *7 | - | - |
| H | desalination-2 | AS12A / carbonate | electric conductivity | 2.701 ± 0.019 (n = 1) *4 | - | - |
*1 Desalination-1 means desalination for Cl-, Br- and I- by the OnGuard II Ag cartridge. Desalination-2 means desalination for Cl-, Br-, I- and SO42- by both the OnGuard II Ag cartridge and the OnGuard II Ba one. *2 Units of each value are mg kg-1. The n value in each parenthesis means the number of the replication for independent measurements by the standard addition method. Each value after "±" means the combined standard uncertainty estimated from both the calibration curve of the standard addition method and the concentration uncertainty of the standard solution. The combined standard uncertainty uc was calculated from the individual standard uncertainties ui using the following equation
*3 DL indicates the detection limit. *4 *5 *6 Both detections with UV absorbance and electric conductivity were simultaneously carried out in every single run. *7 The results were not utilized to determine the certified values of the NMIJ CRMs.
The recovery test for the IC method was also carried out at 225 nm without desalination using the artificial sample. The artificial sample was prepared by adding nitrate of 2.908 mg kg− 1 (equivalent to HSW) to ELSW, whose nitrate mass fraction was up to 0.001 mg kg− 1. The measured concentration, 2.919 mg kg− 1, agreed with the mass fraction in the gravimetric preparation within the expanded uncertainty with a coverage factor of 2 (U(k = 2)), 0.018 mg kg− 1, derived from the calibration curve of the standard addition method. The IC method is operationally simple and reliable, although it is less sensitive than CFA. On the other hand, CFA is highly sensitive, but its operation is much complicated.
To further validate, the existing CRM, MOOS-2, distributed from NRC Canada was analyzed under the condition of the detection at 225 nm with no desalination by using the AS10 column. The certified value of nitrate ions was converted from the amount-of-substance content (nitrite, and “nitrite and nitrate”) in the unit of µmol L− 1 to mass fractions in that of mg kg− 1 using the measured density of 1.0230 g L− 1 (with a relative standard deviation of five measurements, 0.01 %) at 25.0°C. As a result, the measured value of nitrate ions in MOOS-2 (1.335 ± 0.026 mg kg− 1), agreed with the certified value (1.339 ± 0.063 mg kg− 1). The accuracy of the present method was verified at least within the U(k = 2) of MOOS-2.
In case that the seawater samples were desalinated for chloride, bromide and iodide ions, nitrate ions could be determined by using an anion-exchange column with low capacities such as an AS12A column under the conditions of both electric conductivity detection and UV absorbance detection at 210 nm. The measured values from both detection ways agreed with each other (Table 4). Figure 3 shows chromatograms of HSW by using the AS12A column after the desalination with only an OnGuard II Ag cartridge. A large peak of sulfate ions disappeared when the seawater sample was desalinated for sulfate as well as chloride, bromide, and iodide ions; the measured values of nitrate ions were substantially identical for both electric conductivity detection and UV absorbance detection at 210 nm. There was also no significant difference between the measured values in the two desalination modes.
In case that the high-capacity anion-exchange column (IonPac AS23) was used with a carbonate eluent, nitrate ions could be determined when the seawater sample was diluted to 29% by water (Fig. 4). The measured values of nitrate ions (Table 4) well agreed with those under the other measurement conditions mentioned above. Although Ohguni et al. reported direct determination of nitrate in seawater by IC using a high-capacity anion-exchange column [22], the accuracy of their results was at several percent levels even for the relatively high mass fraction of nitrate.
As described above, regardless of the different measurement conditions such as pretreatments of the seawater sample (only dilution or two modes of the desalination), separation conditions (columns and eluents) and detection ways (electric conductivity or UV absorbance), the measured values of nitrate ions agreed with each other within their expanded uncertainties estimated for the standard addition method (Table 4). This fact strongly suggests that any coexisting ions in the seawater samples did not interfere with the measured values for nitrate ions.
Analytical Results Of Nitrite And Phosphate Ions By Ic
Nitrite ions in the seawater samples were measured by using the AS10 column with an eluent of 80 mmol L− 1 NaCl and detecting UV absorbance at 210 nm (Fig. 5) [21]. Because recovery of nitrite ions with the desalination was a little varied, an analysis with no desalination was only carried out. The mass fraction of nitrite ions was (0.0202 ± 0.0030) mg kg− 1 for MSW (in this section, each value following "±" indicates the U(k = 2). The nitrite mass fractions for both HSW and ELSW were less than the detection limit (DL) (0.009 mg kg− 1 in an undiluted sample), which was defined as a mass fraction corresponding to three times the standard deviation for baseline noise near the retention time of nitrite ions. The reason for diluting the seawater samples despite its low mass fractions was that a negative peak appeared around the retention time of 5.5 min with no sample dilution and the peak interfered in the baselines.
Phosphate ions in the seawater samples were measured in the same way as nitrate ions using the AS23 column (Fig. 4). The desalination was not applied for the same reason as nitrite ions using the AS10 column. The mass fractions of phosphate ions in the seawater samples by IC were (0.243 ± 0.030) mg kg− 1, (0.109 ± 0.054) mg kg− 1, and (0.019 ± 0.016) mg kg− 1 for HSW, MSW, and ELSW, respectively. The mass fraction of phosphate ions in ELSW was close to the DL of this method.
Homogeneity Of The Candidate Nmij Crms
According to ISO guide 35 [23], the homogeneities of the seawater samples (candidate NMIJ CRMs) were evaluated based on the analytical results of 10 bottles × 2 times (4 times for nitrate in MSW, HSW and dissolved silica in 3 levels of seawater) by CFA. All the uncertainty due to the homogeneity was shown in Table 5 with that for dissolved silica. Although the homogeneity of the ELSW and a part of HSW are poorer than others due to the extremely low mass fractions of the analyte in those, these amount-of substance components were conclusively decided as indicative values or information as described later section.
Table 5
Standard uncertainty due to the homogeneity for the candidate NMIJ CRMs.
| | HSW / mg kg− 1 | MSW / mg kg− 1 | ELSW / mg kg− 1 |
| NO2− | 0.00020 (15.9%) | 0.00007 (0.36%) | 0.00017 (10.3%) |
| NO3− | 0.0082 (0.30%) | 0.0007 (0.07%) | 0.0001 (6.2%) |
| PO43− | 0.0007 (0.23%) | 0.0002 (0.22%) | 0.0007 (42%) |
| Dissolved silica (as Si) | 0.0017 (0.04%) | 0.0005 (0.06%) | 0.00005 (0.15%) |
| * Each value in parentheses indicates the relative value to the property value. |
Stability Of The Candidate Nmij Crms
The long-term stability of the seawater samples (candidate NMIJ CRMs) was estimated based on the analytical results up to almost 1500 days by CFA. According to ISO guide 35 [23], the trend analysis was performed for the variation of each nutrient mass fraction against the elapsed time, except the components were finally decided as information. As a result, no instability was found for any certified components. The standard uncertainty due to the long-term stability was estimated basically by multiplying the standard deviation of the slope by the time to the expiration date. All the uncertainty due to the long-term stability was shown in Table 6 with that for dissolved silica.
Table 6
Standard uncertainty due to the long-term stability for the candidate NMIJ CRMs.*
| | HSW / mg kg− 1 | MSW / mg kg− 1 | ELSW / mg kg− 1 |
| NO2− | 0.0002 (14%) | 0.0002 (0.90%) | 0.00023 (14%) |
| NO3− | 0.0221 (0.81%) | 0.0098 (1.0%) | 0.00044 (32%) |
| PO43− | 0.0022 (0.75%) | 0.0021 (2.1%) | 0.00052 (34%) |
| Dissolved silica (as Si) | 0.0078 (0.19%) | 0.0045 (0.53%) | 0.00052 (1.5%) |
| * Each value in parentheses indicates the relative value to the property value. |
Comparison of the results by CFA and IC and determination of certified values of the candidate NMIJ CRMs
All analytical results of the candidate NMIJ CRMs were compared. For nitrate ions, all the results listed in Tables 3 and 4 were used. The results for HSW were in good agreement with each other. For MSW, all the results were also in general agreement. For ELSW, although the result by IC was below the DL (0.005 mg kg− 1 as mentioned above), the result by CFA was consistent with the estimation by the DL for IC. The results show CFA and IC can give comparable analysis for those seawater nutrients although sensitivity and the DL level of the analytical methods were different. Although seawater matrix might cause a slight difference between CFA and IC in the case of a low nutrient concentration, that was estimated as uncertainty due to difference among analytical techniques (mentioned below).
The certified values were calculated as an arithmetic means of the quantitative values measured by following three analytical techniques: CFA, IC direct measurement (without desalination) and IC with desalination. The quantitative values by IC with/without desalination were determined as follows, respectively: in case that two simultaneous detections by UV absorbance and electric conductivity were utilized for nitrate (BC, EF and GH in Table 4), a single measured value with a smaller uncertainty was chosen from both detection ways. Then, weighted means were taken for the direct measurement and the measurement with desalination, respectively, using the uncertainty due to the measurements as weight. For evaluating the uncertainty of the certified values, the differences among the analytical techniques, the homogeneity and the stability were taken into consideration. In the case of ELSW, a reliable certified value could not be estimated because of the low mass fraction; then, a property value was not given as a certified value, but just as information of a quantitative value using CFA.
On the nitrite analysis, comparison among the analytical methods could be demonstrated only for MSW due to very low mass fractions of nitrite ions. As a result, the quantitative values by CFA and IC without desalination were in good agreement with each other. A certified value was given to only MSW as an arithmetic means of both results. For nitrite ions in HSW and ELSW, property values were given as information the same as nitrate in ELSW. Due to the relatively low mass fractions of phosphate ions, those uncertainties were extremely high for IC; therefore, the property values of phosphate ions were decided as indicative ones for HSW and MSW, and as information for ELSW, only from the results by CFA.
The certified values, the indicative ones and the information are summarized in Table 7. The certified values for dissolved silica are also shown together in Table 7.
Table 7
Certified values of the three NMIJ CRMs (HSW, MSW and ELSW)*1
| | HSW / mg kg− 1 | MSW / mg kg− 1 | ELSW / mg kg− 1 |
| NO2− | 0.0012 *3 | 0.019 ± 0.002 | 0.0016 *3 |
| NO3− | 2.745 ± 0.050 | 0.942 ± 0.027 | 0.0013 *3 |
| PO43− | 0.288 ± 0.005 *2 | 0.101 ± 0.005 *2 | 0.002 *3 |
| Dissolved silica (as Si) | 4.106 ± 0.043 | 0.837 ± 0.018 | 0.036 ± 0.004 |
| *1 Mass fraction (mg kg− 1). Each value after "±" indicates the expanded uncertainty U(k = 2). |
| *2 Indicative value. |
| *3 Provided as information, that means a quantitative value in time using only one analytical method. Since final uncertainty was large and the reliability as a property value was not sufficient, these values were provided as that. |