The methodology is structured into two distinct phases. The first phase involves the collection of comparative data from human donor specimens through uniaxial compression testing and axial pullout tests of pedicle screws. Subsequently, the second phase encompasses a detailed description of the lattice design, its mechanical characterization, and the comparative evaluation of fully 3D-printed vertebrae.
Human bone characterization
43 vertebrae from 6 human body donors (2 females, 4 males) aged 86.8 ± 7.8 (mean ± standard deviation) years were obtained in fresh and anatomically unfixed condition from levels T7 to L5. All body donors gave their informed and written consent to the donation of their bodies for teaching and research purposes while alive. Being part of the body donor program regulated by the Saxonian Death and Funeral Act of 1994 (third section, paragraph 18 item 8), institutional approval for the use of the post-mortem tissues of human body donors was obtained from the Institute of Anatomy, University of Leipzig (ethical approval No. 129/21-ck). The authors declare that all experiments were conducted according to the principles of the Declaration of Helsinki. All bones were stored fresh frozen at -80°C until further preparation. CT scans (Voltage: 120 kV; slice thickness 1 mm; s. Figure 1, Clinical Imaging) as well as Dual-Energy-X-Ray (DXA) scans were conducted on the frozen whole spines. Before testing all vertebrae were thawed one time to separate each bone from surrounding soft tissues and other bony anatomy. Afterwards the individual bones were freezed again at -80° C until testing.
On the test day, the vertebrae were thawed overnight at 4° C and then a polyaxial pedicle screw (M.U.S.T. Pedicle Screw, Medacta International, Castel San Pietro, Switzerland) was randomly instrumented on one side using the classic trajectory with the freehand technique. This involved pre-drilling with a 2.5 mm diameter and tapping using the respective tools. Each vertebra was aligned in aluminum cylinders using 3D-printed clamps and spacers to align the transverse plane of the vertebra orthogonal to the cylinder axis. The vertebral body was then embedded using a cold-curing cast resin system (RenCast FC 52/53, Huntsman Advanced Materials, Basel, Switzerland). Additional fixing screws were embedded to prevent the casting material from shifting and twisting within the cylinder (s. Figure 1, Specimen Preparation). Now the aluminum cylinder got aligned and secured with screws. This is done by inserting a K-wire into the cannulated screw to visualize the longitudinal axis of the screw. A wire rope was then looped into the screw head and connected to the machine's crosshead using the matching locking screw (s. Figure 1, Pedicle-Screw Pullout). The whole fixture was mounted on a xy-table within the testing machine (Allroundline Z10, Zwick/Roell GmbH & Co. KG, Ulm, Germany), equipped with a 2.5 kN lead cell, to eliminate transverse loads. All pull-out tests were carried out with a preload of 5 N and a test speed of 5 mm/min according to ASTM F543 [23] to determine the maximum pull-out force Fmax. Testing was stopped after 60% of Fmax had been reached.
The embedded vertebra was then extracted from the cylinder and the 3D-printed spacer removed (s. Figure 1, Specimen extraction). The spacer forms a free space above the vertebra which minimizes tool wear and also creates a flat surface below the vertebra, enabling the vertebra to be aligned in a cranio-caudal direction within the stationary drilling machine.
A drill core of Ø6 mm is extracted via a tenon cutter (FAMAG Series 1616, FAMAG Werkzeugfabrik GmbH & Co. KG, Remscheid, Germany) after explanting the pedicle screw. The cylinder was then cut to a 12 mm length using a band saw (EXAKT 310, EXAKT Advanced Technologies GmbH, Norderstedt, Germany) equipped with a specially designed and 3D-printed cutting device [24]. Brass plates were glued to the end faces of each specimen with cyanoacrylate to minimize end-artifacts (s. Figure 1, Compressive Test). Subsequent uniaxial compression testing was also performed using the above-mentioned testing machine, whereby the test load was applied via polished stainless steel test platens. A hysteresis loop between 0.16 MPa and 0.33 MPa is applied as preconditioning. These limits come from pre-tests and are set to avoid irreversible damage to the samples. Afterwards, the samples were tested to a maximal compression of 40%. The determined parameters are Modulus E (maximum slope in the quasilinear range; s. Figure 2, A), compressive stress σy (failure stress at 0.2% offset of the E; s. Figure 2B) and the plateau stress σp (mean value of all stress values in the 20–40% strain range; s. Figure 2C) [24, 25]. Between all steps, the samples were stored in 0.9% saline solution to prevent dehydration. Each vertebra was tested within one day.
The determined mechanical parameters were then correlated with the corresponding HU from the CT data of their respective vertebral hemisphere. HU is determined using segmentation software (Mimics Innovation Suite V.23, Materialise, Leuven, Belgium) in accordance with the preliminary work of [26]. A filled mask is created for each vertebra using the threshold method. The created mask is then thinned using the Erode function so that a new mask is created containing only cancellous bone tissue. Next, this mask is divided into its left and right sides and the average HU of each mask was exported.
Modeling and characterizing the lattice structure
A hexagonal lattice structure replaces the complex architecture of the cancellous bone within the bone. It was created using the computer-aided-design (CAD) software (Rhino 7, Robert McNeel & Associates, Seattle, WA, USA) and consists of equilateral and equiangular hexagons in one plane, the corners of which are in turn connected with vertical bars between the individual planes. The lattice can thus be clearly defined and customized by a single parameter which is the ratio of strut thickness t and strut length L (s. Figure 3).
The average trabecular thickness in lumbar vertebral bodies is approximately .12 mm with an average bone volume fraction of 8.15% [27]. Such small struts cannot be reliably printed by stereolithography (SLA). Initial tests determined a minimum printable strut thickness of t = .4 mm for the SLA-process. This was done by producing cylindrical test specimens with increasing rod thickness, starting at .2 mm, on the SLA printer (Form 3B, Formlabs, Sommerville, MA, USA). The test specimens were then analyzed for printing defects. Since the average trabecular thickness of human cancellous bone in the vertebral body is around .1 mm, the smallest rod diameter of .4 mm that could be reproducibly printed was selected.
Cylindrical specimens with a diameter to length ratio of 1:2 were printed for the mechanical characterization (analogous to the bone specimens; s. Figure 4 right) [15, 24, 28, 29]. The specimen diameter is generally defined to be ten times larger than the largest lattice spacing. This ensures the structural integrity of the sample [29]. Mechanical stability of the structure is therefore ultimately defined by the ratio of t/L (strut diameter/strut length) and is therefore independent of the selected t. Changing t/L means a variation in the solid content and thus a change in the structural behavior. The relationship between t/L and the resulting relative density (solid content) of the lattice structure is shown in (Fig. 3, right). The grid is fused with 1 mm thick plates at the end faces of each cylinder for preventing end artefacts during the compression tests [30]. Additionally, these plates provide an attachment surface for support structures during printing. Six samples of each of the selected t/L ratios were printed from each of two resins (Clear V4 and Tough 2000, Formlabs, Sommerville, MA, USA). The material characteristics of the resins are shown in Table 1.
Table 1
Tensile properties of the used resin materials given by the manufacturer (Formlabs, Sommerville, MA, USA)
|
Material
|
Elastic modulus (GPa)
|
Ultimate tensile strength (MPa)
|
Elongation at break (%)
|
|
Clear V4
|
2.8
|
65
|
6
|
|
Tough 2000
|
2.2
|
46
|
48
|
One sample from each sample group was used for a preliminary test, so that a total of 90 samples were available for characterizing the lattice structure [29]. Characterization of the 3D-printed specimens followed the same procedure as the human bone compressive tests (s. Fig. Figure 4).
The parameterization of the grid structure is based on the mathematical relationships between HU and the mechanical parameters of the human print samples (1), as well as the relationship between t/L and the mechanical properties (2) of the printed grid samples. All correlations are carried out by curve fitting of the experimental data by means of power law relations. Parameterization can be performed via two equations and one unknown for each of the three mechanical parameters (E, σy, σp). The relationship (3) is obtained by inserting and rearranging equations (1) and (2).
$$\:y=a{\bullet\:\left(HU\right)}^{b}$$
1
$$\:y=u{\bullet\:\left(\raisebox{1ex}{$t$}\!\left/\:\!\raisebox{-1ex}{$L$}\right.\right)}^{v}$$
2
$$\:\raisebox{1ex}{$t$}\!\left/\:\!\raisebox{-1ex}{$L$}\right.=\sqrt[v]{\frac{a\bullet\:{HU}^{b}}{u}}$$
3
Since the variables a, b, u and v differ for each mechanical parameter, three different values for t/L are calculated for each printed resin. Averaging these three values gives the corresponding ratio of t/L for the bone model. In this way, all gathered mechanical parameters are included in the model.
3D-printed vertebral models
Three lumbar vertebrae from different donors were selected for replicating whole vertebral bones to cover different bone densities and different lumbar levels. Both the outer and inner contours of the cortical bone were segmented using the threshold method. Care was taken to prevent overlapping contours (see Fig. 5, a). Necessary corrections were made using the ‘Edit Contours’ software tool. The contours were exported from the segmentation software as STL-files and imported into the CAD-software for further processing. First, the models were remeshed with an element size of .8 mm (Fig. 5, b). Model contours were checked again by re-importing the models into the segmentation software. This was followed by generating the parametrized unit cell (Fig. 5, c) via HU of the vertebra (Fig. 5, d). The lattice is then merged with the cortical bone using Boolean operations. Drain holes were integrated into the cover surfaces and the vertebral processes to ensure proper cleaning of the non-polymerized resin.
The resulting 3D models (Fig. 5,e) were then prepared for printing using preprinting software (PreForm®, Formlabs, Sommerville, MA, USA) and sent to the printer. After printing, the models were thoroughly cleaned with isopropanol and post-cured under ultraviolet radiation and temperature according to the manufacturer's instructions. As a pre-test for the pull-out tests, several vertebrae were made from the two previously analyzed resins and instrumented with pedicle screws. The vertebrae made with ClearV4 resin exhibited such brittle behavior that the models failed locally during instrumentation (pre-drilling, tapping, screw insertion). For this reason, whole vertebral models were printed out of Tough2000 V2 resin, which has a higher ultimate strain to avoid these issues (s. Table 1).
One lumbar vertebra with normal bone quality and two vertebrae with osteoporosis from different donors and different levels of the lumbar spine were selected for the final pull-out tests of the 3D-printed bones (s. Table 2). Each of these three human vertebrae was printed five times from Tough2000 V2 resin. The drain holes were sealed with adhesive tape to prevent penetration of the casting resin into the printed vertebrae during embedding.
Table 2
Specifications of the three lumbar vertebrae which were selected for 3D-printing. The selection is justified by ensuring the representation of different bone qualities and different vertebral levels.
|
Donor
|
Level
|
Spongious HU
|
Spongious HU (screw side)
|
|
1
|
L3
|
118
|
93
|
|
3
|
L1
|
303
|
343
|
|
4
|
L4
|
62
|
58
|
Statistical analyses
Descriptive statistics were examined using IBM SPSS Statistics 28 (IBM Corporation, Armonk, NY, USA). Statistical significance is defined with P < .05. Curve estimation based on power law was used to determine relationships between the compressive properties (E, σy, σp) and the HU or the lattice parameter t/L. Linear as well as nonlinear (power, exponential) curve estimations were furthermore applied for analysing the dependence of pullout force and HU. Kruskal-Wallis test for independent samples were used to distinct between the of normal, osteopenic and osteoporotic vertebrae.